KONICA MINOLTA CM-3700A Plus Benchtop Spectrophotometer (Model CM-3700A Plus) is a Color & Light Measurement supplied by Guangzhou Yunke Electronics Co., Ltd. KONICA MINOLTA CM-3700A Plus reference benchtop spectrophotometer, 360-740 nm, 10 nm pitch, Ø152 mm sphere, reflectance and transmittance, inter-instrument ΔE*ab within 0.08.
Brand: KONICA MINOLTA | Category: Color & Light Measurement
Sales Inquiries: 15626018696 · 78403603@qq.com

| Model | CM-3700A Plus |
|---|---|
| Illumination / Measure Welding Systems | Reflection di:8°, de:8° (diffuse illumination, 8° reception), SCI/SCE switchable; transmission di:0°, de:0° |
| Integrating Sphere Size | Ø152 mm (6 in.) |
| Sensor | Silicon photodiode array (38 elements) |
| Spectral Method | Diffraction grating |
| Wavelength Range | 360 ~ 740 nm |
| Wavelength Interval | 10 nm |
| Half Bandwidth | Approx. 14 nm (average) |
| Measurement Range | 0 ~ 200%; resolution 0.01% |
| Light Source | Pulsed xenon lamp |
| Measurement / Illumination Area | Reflection SAV Ø3 mm/Ø5 mm, MAV Ø8 mm/Ø11 mm, LAV Ø16 mm/Ø20 mm, LAV Ø25.4 mm/Ø28 mm; transmission approx. Ø20 mm/Ø25 mm |
| Repeatability | White: within ΔE*ab 0.005; black: within ΔE*ab 0.02 |
| Inter-Instrument Agreement | Within ΔE*ab 0.08 (average of 12 BCRA Series II color tiles, LAV/SCI) |
| Measurement Time | Approx. 2 s for SCI or SCE; approx. 4.5 s for SCI+SCE |
| UV Adjustment | Continuous UV 100% / UV 0% adjustment in 1000 steps |
| Interface | USB 2.0 |
| Power Supply | Dedicated AC adapter |
| Dimensions | Approx. 307 (W) × 271 (H) × 600 (D) mm |
| Weight | Approx. 20.0 kg |
Some specification lines are quoted from the original manufacturer datasheet. Numbers, units and model codes are preserved verbatim; an English datasheet is available on request.
Automotive and transportation, electrical equipment, smartphones, cosmetics, coatings, plastics, building materials and textiles. Typical use covers incoming color inspection, in-line color difference checks, color matching and outgoing appearance verification.
Reflection SAV Ø3 mm/Ø5 mm, MAV Ø8 mm/Ø11 mm, LAV Ø16 mm/Ø20 mm and LAV Ø25.4 mm/Ø28 mm; transmission approx. Ø20 mm/Ø25 mm.
White within ΔE*ab 0.005 and black within ΔE*ab 0.02, as stated in the official specification.
Color and appearance measurement of solid, powder, paste and transparent samples; selection depends on sample size, surface texture and measurement aperture.