| Applicable Standard | IEC 61672-1:2013 Class 2、JIS C 1509-1:2017 Class 2、JIS C 1516:2020 Class 2 |
|---|---|
| Specification | UC-52(Sensitivity Logic Level -33 dB, 1 kHz when Phase to 1 V / Pa) |
| Measuring Range | A Weighted 30–137 dB|C Weighted 36–137 dB|C Weighted Peak 65–140 dB |
| Noise | A Weighted ≤24 dB(Typical 22 dB)|C Weighted ≤30 dB(Typical 28 dB) |
| Measuring Frequency Range | 20 Hz–8 kHz |
| Frequency Weighted | A、C |
| Time Weighted | F(), S() |
| Measuring Item | Lp、Leq、LE、Lmax、Lpeak |
| Sampling | Lp/Lmax/Lpeak:20.8 μs(48 kHz);Leq/LE:1 s |
| Display | 1.5 in Single with Unit EL, Resolution 128×128; Value New 1 s, records Diagram 100 ms |
| Memory | internal Part Memory Device up to 200 groups |
| Calibration | supports NC-75 / NC-74 Calibration Device Automatic Calibration, can Management Management up to 999 times Calibration |
| Interface | USB Type-C(Electric is can Shift Magnetic Disk, Direct CSV) |
| Power Supply | 2 cells AA Battery |
| Battery Battery Life | Alkaline Battery Approx. 9 hours (23 ℃, Measurement ) |
| Dimensions | Approx. 200(H)× 50(W)× 34(D)mm |
| Weight | Approx. 160 g(incl. battery) |
| Operating Environment | -10 ℃~50 ℃, 10%~90%RH(non-condensing) |
Some specification lines are quoted from the original manufacturer datasheet. Numbers, units and model codes are preserved verbatim; an English datasheet is available on request.
RION Japan RION NL-28 portable sound scale 2: 20 Hz-8 kHz, single hand holding, 2 Section AA Battery Extension of approximately 9 hours to support USB-C export CSV
Q: Can 1/3 be observed multiple frequency?. Reference: NL-28
A: No, NL-28 is a sound scale, providing only a total value measurement; for spectrum analysis, please select NA-28
Q: How do you get the data out?
A: Connect computers through USB Type-C, where internal memory is recognized as a mobile disk, directly copying CSV files
Q: Must the calibration be bought?
A: It is recommended that NC-75 sound calibration be equipped for pre- and post-measure sensitivity verification to ensure that data are traceable
Q: Are there matching and technical options available?
A: RION provides a complete system of matching and fittings, with specific configurations that can be identified by contacting Guangzhou Yunke Electronics Co., Ltd