The SEFAR (Switzerland) TENSOCHECK 200 is an electronic digital instrument for measuring screen mesh tension, commonly used for incoming inspection of SMT stencils and screen-printing meshes and for confirming tension after stretching. It measures the static tension of the stretched mesh, in N/cm. The industry-standard judgement method is the five-point method recommended by ESMA. Below, in the order of actual operation, we cover preparation, point layout, reading, judgement and common mistakes in one pass.
I. First, confirm what this instrument measures and what it does not
- Can measure: static tension of polyester mesh, stainless steel mesh and other fabrics after stretching on a frame; SMT laser stencils, screen-printing meshes and textile printing screens are all covered.
- Cannot measure: dynamic tension of running wire, yarn or optical fibre — that is the domain of wheel-type tension meters (such as the ZEF-100 digital tension meter), whose principle and probe structure are completely different.
- It measures point values: the instrument gives the local tension at the probe position, so measurements must follow a fixed point layout; a single reading cannot represent the whole screen.
II. Structure and operating interface
- Probe: the measuring foot presses vertically onto the mesh; an internal force sensor outputs the tension value.
- Display: colour LCD with 10 mm digit height, reading direction aligned with the measurement direction, so bending over to read does not easily cause mistakes.
- Glass reference plate: the supplied 80 × 125 × 7 mm glass plate, used for daily zero check without any external equipment.
- WiFi: the device provides its own hotspot; connect a phone or computer and open a browser to access 192.168.1.1 for remote reading, logging and CSV export.
- Power: 3.7 V / 7.2 Wh lithium-ion battery, USB-C charging; adjustable auto power-off time.
III. Main specifications (official data)
| Item | Specification |
|---|---|
| Measuring range | 4.0 – 50.0 N/cm |
| Resolution | 0.2 N/cm(≤ 29.8 N/cm);0.5 N/cm(30 – 50 N/cm) |
| Dimensions | 80 × 95 × 115 mm |
| Weight | 1.01 kg (including battery) |
| Power supply | 3.7 V / 7.2 Wh lithium battery, USB-C charging |
| Data | WiFi remote operation + CSV recording and export |
| Units | Switchable between N/cm and mm |
Permissible error is graded by measuring range; check the applicable band when judging, rather than applying a single figure throughout:
| Measuring interval (N/cm) | Permissible error |
|---|---|
| 4.0 – 9.8 | ± 0.2 N/cm |
| 10 – 19.8 | ± 0.4 N/cm |
| 20 – 29.8 | ± 0.6 N/cm |
| 30 – 39.5 | ± 1.0 N/cm |
| 40 – 50.0 | ± 1.5 N/cm |
IV. Four preparations before measuring
- Wait for tension to stabilise: tension in a freshly stretched screen has not stabilised; measuring immediately gives high readings that drift. Follow the in-house process card for resting time, do not rely on feel.
- Clean the mesh: remove emulsion residue, dust and oil; no foreign matter between probe and mesh.
- Check battery and self-test: wait for the self-test to complete after power-on before measuring; low battery directly affects reading stability.
- Perform a zero check: see Section VI. This step is the most often skipped and the most common cause of systematic bias across an entire batch of data.
V. Five-point measurement procedure (recommended by ESMA)
Screen tension cannot be judged from the centre point alone.Divide the length and width of the screen into thirds to form a nine-square grid: 1 centre point plus the 4 intersections of the dividing lines, 5 points in total.These 5 points cover the effective printing area and reflect uniformity across the whole screen.
- Point layout: measure the first point at the centre of the screen; the remaining 4 points sit at the intersections of the grid lines.
- Orientation: all points are read along the squeegee stroke direction, regardless of mesh angle. If orientations differ, the readings are not comparable.
- Placement: hold the probe vertically and place it gently on the mesh; do not press down or tilt. Hand pressure adds directly to the reading.
- Reading: read once the value stabilises; repeat 2–3 times at each point and average, discarding obviously erratic single readings.
- Recording: record all 5 values. For batch inspection, connect over WiFi and export CSV to avoid transcription errors.
- Judgement:Deviation between the 5 points should not exceed ± 1 – 2 N/cm; screens for the same multi-colour print set must also fall within the same tolerance band.
- Handling anomalies: if one point is clearly low, first check that area for slack mesh, adhesive failure or damage; if all 5 points are low, the stretching tension is insufficient and the screen must be re-stretched, rather than adjusting printing parameters to compensate.
VI. Zero check and accuracy verification
- Daily zeroing: place the instrument gently on the glass reference plate (probe unloaded); the reading should be 0.00. If the deviation exceeds the permissible error for that band (about ± 0.2 N/cm in the low range), zero it with the panel ZERO function; if it still drifts after zeroing, stop using it and send it for calibration.
- Frequency: recommended before each measurement or at the start of each shift, especially after transport or temperature changes.
- Periodic verification: an optional SEFAR calibration control plate (122 × 80 × 8 mm, supplied with an SCS calibration certificate) can be used for periodic accuracy checks; the control plate itself also requires periodic recalibration by an accredited metrology body under ISO 9001.
- Send for service only when out of tolerance: this is the whole point of routine verification — if readings stay within tolerance there is no need to send the whole instrument for calibration, avoiding unnecessary freight and calibration cost.
VII. Data logging and traceability
Once a phone or computer is connected to the device hotspot, open 192.168.1.1 in a browser for remote control: read live values, log measuring points and export historical data as CSV. For batch incoming inspection, use one file per screen with the screen number and date in the filename; later traceability is far faster than paper records.
VIII. Common mistakes and corrections
| Mistake | Consequence | Correct practice |
|---|---|---|
| Pressing down or tilting the probe | High readings, poor repeatability | Place vertically and gently; the hand only steadies the instrument |
| Measuring close to the frame or glued edge | Edge tension reads significantly high | Follow the five-point layout, avoid the frame and glued area |
| Inconsistent orientation between points | Data not comparable, uniformity misjudged | Read every point along the squeegee stroke direction |
| Skipping the zero check | Systematic bias across the whole batch | Zero on the glass reference plate each shift |
| Mixing N/cm and mm | Wrong acceptance criteria | Judge consistently in N/cm |
| Measuring immediately after stretching | High readings with continuous drift | Re-measure per the process card once tension stabilises |
IX. Maintenance notes
- Keep the glass reference plate clean and scratch-free; scratches invalidate the zero check.
- Avoid drops and over-range pressing; protect the probe from load during transport.
- Top up the battery periodically during long storage; lithium batteries should not be left discharged.
- Store in a dry environment, away from high temperatures and strong magnetic fields.
- Send for periodic calibration as required by the quality system and retain calibration records.
X. Frequently asked questions
Q: Can it measure above 50 N/cm?
The measuring range is 4.0 – 50.0 N/cm. Values displayed outside this range carry no accuracy guarantee and cannot be used for judgement.
Q: Can it measure both polyester and stainless steel mesh?
The official positioning covers all types of mesh and frames. Specific tension targets depend on mesh specification, mesh count and in-house process; the instrument is responsible only for an accurate reading.
Q: A 3 N/cm spread across the 5 points — is that a fail?
The acceptance line is ± 1 – 2 N/cm. If exceeded, first check for local slack or adhesive failure before deciding to re-stretch.
Q: The reading keeps fluctuating — what should I do?
Check in order: whether the mesh is still stabilising, whether the probe is sitting flat, whether the battery is low, and whether there is a vibration source nearby.
Q: Can this measure wire tension?
No. Dynamic tension of wire and yarn requires a wheel-type tension meter; the two are not interchangeable.
XI. Sources
- SEFAR Tensocheck 200 official product literature: measuring range, resolution, graded permissible error, dimensions and weight, battery and WiFi specifications, glass reference plate dimensions.
- ESMA (European Screen Printing Manufacturers Association) five-point screen tension measurement method: point layout, measurement orientation and tolerance judgement.
- Product page on this site: SEFAR Tensocheck 200 Mesh Tension Meter; for more models see the Tension Testers category.