The SEFAR (Switzerland) TENSOCHECK 200 is an electronic digital instrument for measuring screen mesh tension, commonly used for incoming inspection of SMT stencils and screen-printing meshes and for confirming tension after stretching. It measures the static tension of the stretched mesh, in N/cm. The industry-standard judgement method is the five-point method recommended by ESMA. Below, in the order of actual operation, we cover preparation, point layout, reading, judgement and common mistakes in one pass.

I. First, confirm what this instrument measures and what it does not

II. Structure and operating interface

III. Main specifications (official data)

ItemSpecification
Measuring range4.0 – 50.0 N/cm
Resolution0.2 N/cm(≤ 29.8 N/cm);0.5 N/cm(30 – 50 N/cm)
Dimensions80 × 95 × 115 mm
Weight1.01 kg (including battery)
Power supply3.7 V / 7.2 Wh lithium battery, USB-C charging
DataWiFi remote operation + CSV recording and export
UnitsSwitchable between N/cm and mm

Permissible error is graded by measuring range; check the applicable band when judging, rather than applying a single figure throughout:

Measuring interval (N/cm)Permissible error
4.0 – 9.8± 0.2 N/cm
10 – 19.8± 0.4 N/cm
20 – 29.8± 0.6 N/cm
30 – 39.5± 1.0 N/cm
40 – 50.0± 1.5 N/cm

IV. Four preparations before measuring

  1. Wait for tension to stabilise: tension in a freshly stretched screen has not stabilised; measuring immediately gives high readings that drift. Follow the in-house process card for resting time, do not rely on feel.
  2. Clean the mesh: remove emulsion residue, dust and oil; no foreign matter between probe and mesh.
  3. Check battery and self-test: wait for the self-test to complete after power-on before measuring; low battery directly affects reading stability.
  4. Perform a zero check: see Section VI. This step is the most often skipped and the most common cause of systematic bias across an entire batch of data.

V. Five-point measurement procedure (recommended by ESMA)

Screen tension cannot be judged from the centre point alone.Divide the length and width of the screen into thirds to form a nine-square grid: 1 centre point plus the 4 intersections of the dividing lines, 5 points in total.These 5 points cover the effective printing area and reflect uniformity across the whole screen.

  1. Point layout: measure the first point at the centre of the screen; the remaining 4 points sit at the intersections of the grid lines.
  2. Orientation: all points are read along the squeegee stroke direction, regardless of mesh angle. If orientations differ, the readings are not comparable.
  3. Placement: hold the probe vertically and place it gently on the mesh; do not press down or tilt. Hand pressure adds directly to the reading.
  4. Reading: read once the value stabilises; repeat 2–3 times at each point and average, discarding obviously erratic single readings.
  5. Recording: record all 5 values. For batch inspection, connect over WiFi and export CSV to avoid transcription errors.
  6. Judgement:Deviation between the 5 points should not exceed ± 1 – 2 N/cm; screens for the same multi-colour print set must also fall within the same tolerance band.
  7. Handling anomalies: if one point is clearly low, first check that area for slack mesh, adhesive failure or damage; if all 5 points are low, the stretching tension is insufficient and the screen must be re-stretched, rather than adjusting printing parameters to compensate.

VI. Zero check and accuracy verification

VII. Data logging and traceability

Once a phone or computer is connected to the device hotspot, open 192.168.1.1 in a browser for remote control: read live values, log measuring points and export historical data as CSV. For batch incoming inspection, use one file per screen with the screen number and date in the filename; later traceability is far faster than paper records.

VIII. Common mistakes and corrections

MistakeConsequenceCorrect practice
Pressing down or tilting the probeHigh readings, poor repeatabilityPlace vertically and gently; the hand only steadies the instrument
Measuring close to the frame or glued edgeEdge tension reads significantly highFollow the five-point layout, avoid the frame and glued area
Inconsistent orientation between pointsData not comparable, uniformity misjudgedRead every point along the squeegee stroke direction
Skipping the zero checkSystematic bias across the whole batchZero on the glass reference plate each shift
Mixing N/cm and mmWrong acceptance criteriaJudge consistently in N/cm
Measuring immediately after stretchingHigh readings with continuous driftRe-measure per the process card once tension stabilises

IX. Maintenance notes

X. Frequently asked questions

Q: Can it measure above 50 N/cm?
The measuring range is 4.0 – 50.0 N/cm. Values displayed outside this range carry no accuracy guarantee and cannot be used for judgement.

Q: Can it measure both polyester and stainless steel mesh?
The official positioning covers all types of mesh and frames. Specific tension targets depend on mesh specification, mesh count and in-house process; the instrument is responsible only for an accurate reading.

Q: A 3 N/cm spread across the 5 points — is that a fail?
The acceptance line is ± 1 – 2 N/cm. If exceeded, first check for local slack or adhesive failure before deciding to re-stretch.

Q: The reading keeps fluctuating — what should I do?
Check in order: whether the mesh is still stabilising, whether the probe is sitting flat, whether the battery is low, and whether there is a vibration source nearby.

Q: Can this measure wire tension?
No. Dynamic tension of wire and yarn requires a wheel-type tension meter; the two are not interchangeable.