Mitutoyo 513-477-10E Measuring Tools (Model 513-477-10E) is a Mitutoyo Measuring Tools supplied by Guangzhou Yunke Electronics Co., Ltd. Mitutoyo 513-477-10E Measuring Tools — Mitutoyo Measuring Tools · Yunke Electronics. Request a quotation and datasheet.
Brand: Mitutoyo | Category: Mitutoyo Measuring Tools
Sales Inquiries: 15626018696 · 78403603@qq.com

| Order No. | 513-477-10E |
|---|---|
| Series | Mitutoyo Mitutoyo 513 Series Rod Gauge (TI, Lever-Type Test Indicators / Rod Dial Indicator) |
| Structure Type | Horizontal Type · Standard / Non- Switching Type |
| Catalogue Model | TI-123ERX |
| Graduation | 0.01 mm |
| Measuring Range | 1.0 mm |
| Scale | 0-50-0 |
| Indication Error MPE (Full Measuring Range) | 10 µm |
| Indication Error MPE (1 Travel per Revolution) | Official not (—) |
| Indication Error MPE (1 / 10 Travel per Revolution) | 5 µm |
| Hysteresis | 5 µm |
| Repeatability | 3 µm |
| Measuring Force | 0.2 N or less |
| Body Mass | 45 g(excluding with Groove Shaft sets) |
| Product Features | Stylus / Length Stylus Type |
| Standard Stylus | ø2 mm Hardness Stylus ( Stylus Unit Type with Head) |
| Standard Supplied Accessories | ø8 mm Shaft sets (with Groove), Clamp Ring, ø2 mm Hardness Stylus |
| Set Installed Configuration | sets Installed (E), Official Catalogue not Series Same Specifications A / T sets Installed |
| Bearing / Movement | Bearing, Non- Device Structure Measurement Direction Automatic Rotation |
| Crystal and Sealing | Flat Protection Crystal with Protection Coating; Gauge Revolution and Crystal a Body Structure + O Type Revolution, Control and Oil |
| Accuracy Standard | Compliant with JIS B 7533:2015; Horizontal Type / Type / Direct Type per Pointer Surface toward above Inspection, Flat Type per Pointer Surface Direct Inspection |
| Operating Notes | Replacement Stylus Rear Use Reference Device Calibration; Error Use Stylus Guide Measurement Error; Pointer and Hardness Stylus is Magnetic Property, Part Division internal Part Zero pcs incl. Magnetic Property Material |
| Specification Source | Mitutoyo Product Catalogue 2025-2026 No. 07-62 / 07-63 ; Mitutoyo in Official 513 Series Specifications PDF |
Some specification lines are quoted from the original manufacturer datasheet. Numbers, units and model codes are preserved verbatim; an English datasheet is available on request.
Mitutoyo 513 Series TI is a mechanical lever-type test indicator (lever-type dial indicator) in which the swing of the contact point drives the dial pointer, designed for narrow, deep and concave positions that an ordinary plunger-type dial indicator cannot reach. This model 513-477-10E belongs to theHorizontal typeconstruction, graduation 0.01 mm, measuring range 1.0 mm, scale 0-50-0, MPE over the full measuring range 10 µm, measuring force 0.2 N or less. The dial is parallel to the contact point axis, the most widely used standard construction; it can be mounted from the front, top or rear, for machine tool alignment, flatness and runout checks.
For accuracy, the official catalog lists the error terms separately: MPE over 1 revolution is not specified (the pointer travels less than one full revolution at this range), MPE over 1/10 revolution 5 µm, hysteresis error 5 µm, repeatability 3 µm, and body mass 45 g (excluding the stem with dovetail). The whole series uses jeweled bearings and a clutchless mechanism: the measuring direction reverses automatically when the contact point moves up or down, with no selector lever to operate. The bezel and crystal are integrated and sealed with an O-ring to limit cutting fluid and oil ingress from the front of the dial and through gaps (not a waterproof specification).
This model is positioned asRuby contact point / long contact point type. A long contact point with a ruby ball tip balances deep-cavity reach and wear resistance. For selection, first choose the accuracy class by graduation (0.01 / 0.002 / 0.001 mm), then choose the order number by measuring range and body type. The horizontal group contains 16 models; differences are concentrated in measuring range, graduation, contact point length and contact point material. The comparison table below lists every order number in the group for on-site comparison. The photograph shows the same series and the same external form from the official Mitutoyo catalog; contact point length and accessories are as stated in the technical specifications and the official dimensional drawing on this page.
| Order No. | 513-477-10E |
| Series | Mitutoyo 513 Series Lever-Type Test Indicators (TI, lever-type dial indicators) |
| Body type | Horizontal type · standard / non-switchable |
| Catalog order number | TI-123ERX |
| Graduation | 0.01 mm |
| Measuring range | 1.0 mm |
| Scale specification | 0-50-0 |
| Indication error MPE (full measuring range) | 10 µm |
| Indication error MPE (1 revolution travel) | Not specified by the manufacturer (—) |
| Indication error MPE (1/10 revolution travel) | 5 µm |
| Hysteresis error | 5 µm |
| Repeatability | 3 µm |
| Measuring force | 0.2 N or less |
| Body mass | 45 g (excluding the stem with dovetail) |
| Product features | Ruby contact point / long contact point type |
| Standard contact point | ø2 mm carbide contact point (ruby ball tip on ruby contact point models) |
| Standard accessories | ø8 mm stem (with dovetail), knurled clamp ring, ø2 mm carbide contact point |
| Set configuration | Basic set (E); no equivalent A / T sets are listed in the official catalog |
| Bearings / movement | Jeweled bearings; clutchless mechanism reverses the measuring direction automatically |
| Crystal and sealing | Flat anti-glare crystal with a scratch-resistant coating; integrated bezel and crystal with an O-ring limits cutting fluid and oil ingress |
| Accuracy standard | Conforms to JIS B 7533:2015; horizontal / tilted / vertical types are inspected with the dial facing up, and parallel types with the dial vertical |
| Handling precautions | Recalibrate against a reference standard after changing the contact point; using the wrong contact point causes measurement errors; the pointer and carbide contact point are weakly magnetic, while some internal parts still contain magnetic materials |
| Order No. | Graduation (mm) | Measuring range (mm) | Scale specification | Full-range MPE (µm) | Hysteresis error (µm) | Repeatability (µm) | Measuring force |
|---|---|---|---|---|---|---|---|
| 513-424-10E | 0.01 | 0.5 | 0-25-0 | 6 | 4 | 3 | 0.3 N or less |
| 513-478-10E | 0.01 | 0.5 | 0-25-0 | 6 | 4 | 3 | 0.3 N or less |
| 513-466-10E | 0.01 | 0.5 | 0-25-0 | 6 | 4 | 3 | 0.3 N or less |
| 513-414-10E | 0.01 | 0.5 | 0-25-0 | 10 | 5 | 3 | 0.2 N or less |
| 513-426-10E | 0.01 | 1.5 | 0-25-0 | 16 | 5 | 3 | 0.4 N or less |
| 513-404-10E | 0.01 | 0.8 | 0-40-0 | 9 | 4 | 3 | 0.3 N or less |
| 513-474-10E | 0.01 | 0.8 | 0-40-0 | 9 | 4 | 3 | 0.3 N or less |
| 513-464-10E | 0.01 | 0.8 | 0-40-0 | 9 | 4 | 3 | 0.3 N or less |
| 513-415-10E | 0.01 | 1.0 | 0-50-0 | 10 | 5 | 3 | 0.2 N or less |
| 513-477-10E | 0.01 | 1.0 | 0-50-0 | 10 | 5 | 3 | 0.2 N or less |
| 513-405-10E | 0.002 | 0.2 | 0-100-0 | 4 | 3 | 1 | 0.3 N or less |
| 513-475-10E | 0.002 | 0.2 | 0-100-0 | 4 | 3 | 1 | 0.3 N or less |
| 513-465-10E | 0.002 | 0.2 | 0-100-0 | 4 | 3 | 1 | 0.3 N or less |
| 513-425-10E | 0.002 | 0.6 | 0-100-0 | 7 | 4 | 1 | 0.4 N or less |
| 513-401-10E | 0.001 | 0.14 | 0-70-0 | 4 | 3 | 1 | 0.3 N or less |
| 513-471-10E | 0.001 | 0.14 | 0-70-0 | 4 | 3 | 1 | 0.3 N or less |
Note: MPE is the maximum permissible error, guaranteed in the inspection orientation specified by the manufacturer (dial facing up for horizontal / tilted / vertical types, dial vertical for parallel types); "—" means the pointer travels less than one full revolution at this range and the manufacturer does not specify this item. Mass excludes the stem with dovetail.
Specifications on this page are taken from the Mitutoyo Product Catalog 2025-2026, pages 07-62 / 07-63, cross-checked against the official Mitutoyo China specification PDFs "513 Series — Horizontal Type TI" and "513 Series — Tilted / Vertical / Parallel Types" and the Mitutoyo Product Catalog 2020-2021, page F-65. This is a quotation-only catalogue; no prices are listed.
For a quotation, lead time, an English datasheet or an optional calibration certificate, please contact our engineers.