KONICA MINOLTA CA-500 Series Color Analyzer is a Color & Light Measurement supplied by Guangzhou Yunke Electronics Co., Ltd. KONICA MINOLTA CA-500 series color analyzer (CA-527/CA-510), Ø27 mm or Ø10 mm aperture, from 0.0001 cd/m², Flicker, JEITA, VESA, FMA and VRR flicker measurement.
Brand: KONICA MINOLTA | Category: Color & Light Measurement
Sales Inquiries: 15626018696 · 78403603@qq.com

| Model | CA-527 / CA-510 |
|---|---|
| Measurement Aperture | CA-527:Ø27 mm; CA-510:Ø10 mm |
| Receiving Angle | CA-527:±8.5°; CA-510:±15° |
| Measurement Distance for Guaranteed Accuracy | 30 ± 5 mm |
| Luminance Range for Guaranteed Accuracy | CA-527:0.0001 ~ 10,000 cd/m²; CA-510:0.0002 ~ 20,000 cd/m² |
| Luminance Accuracy (White) | CA-527:>0.0001 cd/m² ±9%, >0.0005 ±3%, >0.001 ±2%, >0.01 ±1.5%, >0.1 ±1.5%; CA-510:>0.0002 ±9%, >0.0005 ±3.5%, >0.001 ±3%, >0.01 ±2%, >0.1 ±1.5% |
| Luminance Repeatability (2σ) | CA-527:>0.0001 cd/m² 10%, >0.0005 2%, >0.001 1%, >0.01 0.30%, >0.1 0.12%, >1 0.10%; CA-510:>0.0002 10%, >0.0005 3%, >0.001 2%, >0.01 0.35%, >0.1 0.12%, >1 0.10% |
| Luminance Range for Chromaticity Measurement | CA-527:0.001 ~ 10,000 cd/m²; CA-510:0.002 ~ 20,000 cd/m² |
| Chromaticity Accuracy (White) | CA-527:>0.001 cd/m² ±0.003, >0.01 ±0.002; CA-510:>0.003 cd/m² ±0.003, >0.01 ±0.002 |
| Chromaticity Repeatability (2σ) | CA-527:>0.001 cd/m² 0.0030, >0.01 0.0009, >0.1 0.0004, >1 0.0002; CA-510:>0.003 0.0030, >0.01 0.0012, >0.1 0.0005, >1 0.0003 |
| Flicker (Contrast Method) | Flicker: luminance 0.5 ~ 10,000 cd/m² (CA-527) / 1 ~ 20,000 cd/m² (CA-510), 0.25 ~ 65 Hz; accuracy ±0.3% (30/60 Hz, AC/DC 10% sine wave); repeatability 0.3% |
| Flicker (JEITA) | Flicker: luminance 0.5 ~ 10,000 cd/m² (CA-527) / 1 ~ 20,000 cd/m² (CA-510), 0.42 ~ 65 Hz; accuracy ±0.35 dB (30 Hz, AC/DC 4%/1.2% sine wave); repeatability 0.1 dB / 0.3 dB |
| Waveform Measurement | Flicker sampling frequency 200 kHz adjustable (luminance 0.1 ~ 10,000 cd/m²); XYZ sampling frequency 3 kHz variable (luminance 0.1 ~ 10,000 cd/m²) |
| VRR Flicker | Flicker: sampling frequency 200 kHz adjustable, target 0.01 ~ 100 Hz; accuracy ±0.3% (1 ~ 120 Hz, AC/DC 10% sine wave); repeatability 0.3% |
| Measurement Speed | Lvxy AUTO: 1 reading/s (>0.0001 cd/m²), 5 readings/s (>0.012 cd/m²), 20 readings/s (>0.15 cd/m²); flicker (contrast method) 20 readings/s |
| Measurement Target (Vertical Sync Frequency) | 0.5 ~ 240 Hz (luminance and chromaticity) |
| Interface | USB2.0、RS-232C |
| Power Supply | DC 5 V (via USB bus power or RS communication connector input) |
| Operating Temperature / Humidity Range | 10 ~ 35°C, relative humidity 85% or less, no condensation |
| Storage Temperature / Humidity Range | 0 ~ 45°C, relative humidity 85% or less (at 35°C), no condensation |
| Dimensions | 52 × 52 × 272 mm (excluding mounting base) |
| Weight | 710 g (including mounting base) |
OLED, Micro LED and Micro OLED displays, AR/VR wearables, medical displays and micro projectors. Typical use covers luminance, chromaticity, gamma and flicker evaluation for new display technologies.
Ø27 mm on the CA-527 and Ø10 mm on the CA-510, with a guaranteed-accuracy measurement distance of 30 ± 5 mm.
Luminance 10% at 0.0001 cd/m², 2% at 0.0005 cd/m² and 1% at 0.001 cd/m² on the CA-527, improving to 0.10% above 1 cd/m².
High-resolution and low-luminance displays, including OLED, Micro LED and Micro OLED panels for AR/VR and medical applications.