Technical Specifications

ModelCA-410 series (probes CA-VP427A / CA-P427)
Measurement ApertureØ27 mm
Measurement Distance for Guaranteed Accuracy30 ± 10 mm
Luminance Range for Guaranteed Accuracy0.001 ~ 5,000 cd/m²
Luminance Accuracy (White)>0.0005 cd/m² ±2%, >0.001 ±2%, >0.01 ±2%, >0.1 ±2%, >1 ±2%, >10 ±2%, >100 ±2%
Luminance Repeatability (2σ)>0.0005 cd/m² 5%, >0.001 2%, >0.01 0.5%, >0.1 0.1%, >1 0.1%, >10 0.1%, >100 0.1%
Luminance Range for Chromaticity Measurement0.001 ~ 5,000 cd/m²
Chromaticity Accuracy (White)>0.001 cd/m² ±0.003; 100 cd/m² monochrome ±0.003
Chromaticity Repeatability (2σ)>0.001 cd/m² 0.0015, >0.01 0.0005, >0.1 0.0003, >1 0.0002, >10 0.0002, >100 0.0002
Flicker (Contrast Method)Flicker: luminance 0.5 ~ 5,000 cd/m², 0.25 ~ 65 Hz; accuracy ±0.3% (30/60 Hz, AC/DC 10% sine wave); repeatability 0.3%
Flicker (JEITA)Flicker: luminance 0.5 ~ 5,000 cd/m², 0.42 ~ 65 Hz; accuracy ±0.35 dB (30 Hz, AC/DC 4%/1.2% sine wave); repeatability 0.1 dB / 0.3 dB
Measurement Target (Vertical Sync Frequency)0.5 ~ 240 Hz (luminance, chromaticity); 0.5 ~ 130 Hz (flicker)
InterfaceUSB2.0、RS-232C
Power SupplyDC 5 V (via USB bus power or RS communication connector input)
Operating Temperature / Humidity Range10 ~ 35°C, relative humidity 85% or less, no condensation
Storage Temperature / Humidity Range0 ~ 45°C, relative humidity 85% or less (at 35°C), no condensation
DimensionsCA-VP427A:47 × 47 × 190.5 mm; CA-P427:42 × 42 × 139.7 mm
WeightCA-VP427A: 510 g (including mounting base); CA-P427: 270 g (including mounting base)

Key Advantages

Applications

LCD monitors, televisions and projectors, automotive displays and lighting devices. Typical use covers luminance and chromaticity inspection on production lines.

FAQ

Which probes are available?

CA-VP427A and CA-P427, both with a Ø27 mm measurement aperture.

What repeatability does it achieve?

Luminance 5% at 0.0005 cd/m², 2% at 0.001 cd/m² and 0.1% above 0.1 cd/m².

Which samples is it intended for?

Displays, televisions and lighting devices requiring stable luminance and chromaticity measurement in volume production.

Related Products

KONICA MINOLTA CA-2500 2D Color Analyzer
Color & Light Measurement

KONICA MINOLTA CA-2500 2D Color Analyzer

CA-2500
└ Color Analyzers
KONICA MINOLTA SIG-400 Imaging Color Measurement System
Color & Light Measurement

KONICA MINOLTA SIG-400 Imaging Color Measurement System

SIG-400
└ Color Analyzers
KONICA MINOLTA CA-500 Series Color Analyzer
Color & Light Measurement

KONICA MINOLTA CA-500 Series Color Analyzer

└ Color Analyzers
KONICA MINOLTA T-10A Series Illuminance Meter
Color & Light Measurement

KONICA MINOLTA T-10A Series Illuminance Meter

└ Illuminance Meters
KONICA MINOLTA CL-70F Color Rendering Illuminance Meter
Color & Light Measurement

KONICA MINOLTA CL-70F Color Rendering Illuminance Meter

CL-70F
└ Illuminance Meters
KONICA MINOLTA CL-200A Color Illuminance Meter
Color & Light Measurement

KONICA MINOLTA CL-200A Color Illuminance Meter

CL-200A
└ Illuminance Meters