IMADA force gauges are among the most common force-measuring tools in factories, covering insertion/extraction force, button feel, terminal pull-off and spring pressure. The IMADA force gauges available on this site fall into two families: mechanical dial (PS / PSM series) and digital (ZTS / ZTA series), and their operating logic differs considerably: mechanical models use a tare ring and a peak switch, digital models use ZERO / PEAK buttons. Below, in the order of actual operation, we cover preparation, zeroing, reading, mounting on a test stand, judgement and common mistakes in one pass.
I. First, confirm what this instrument measures and what it does not
- Can measure: static force in both tension and compression — insertion/extraction force, button force, spring pressure, terminal pull-off force, peel force, snap-fit force and similar.
- Cannot measure: impact loads and destructive testing. The IMADA official manual explicitly states that "mechanical force gauges are not recommended for destructive testing"; impact loading directly damages the mechanism.
- It measures axial force: force must be applied along the measuring axis. Side loads introduce significant error and can bend the measuring shaft.
- Handheld or stand-mounted: handheld use is acceptable for quick screening at low capacities; any measurement requiring repeatability and traceability should be done on a test stand (see Section VII).
II. Mechanical vs digital: structural differences determine operating differences
- Mechanical (PS / PSM): spring mechanism plus pointer scale, no power required. Readings are live value / peak hold (switched by a switch on the side of the body); zeroing is done by rotating the tare ring, which also compensates for the weight of attachments.
- Digital (ZTS / ZTA): electronic force sensor plus digital display with 2000 Hz high-speed sampling, capturing peaks and small force changes; ZTS / ZTA use a separate sensor design, with the sensor separate from the display unit for equipment integration or special mounting conditions.
- Common points: both use an M6 thread on the measuring shaft, and attachments are fitted by hand (the official instructions require no tools); both measure in tension and compression.
III. Four preparations before measuring
- Choose the right capacity: the force to be measured should fall within 20%–80% of full scale. Continuous use near full scale accelerates mechanism fatigue, while an oversized capacity wastes resolution.
- Fit the attachment: hand-tighten the required attachment (flat tip, hook, round tip, etc.) onto the measuring shaft; do not use a spanner — the thread is M6 and tool tightening can damage the thread or seize the attachment.
- Zero: on mechanical models rotate the tare ring to bring the pointer to 0, which also removes the influence of attachment weight and mounting orientation; on digital models press ZERO with no load applied.
- Confirm direction: tension tests use the tension shaft and compression tests the compression shaft; do not mix them (see Section IX).
IV. Mechanical operating procedure (PS / PSM series)
- 1. Set the peak switch to OFF: before starting (mechanical models have no power switch, but the state must be reset), set the Peak Switch to OFF so the head shows live values.
- 2. Hand-tighten the attachment: screw the selected attachment onto the measuring shaft and confirm it is secure.
- 3. Zero with the tare ring: rotate the Tare Ring until the pointer aligns with the 0 graduation. Re-zero after every attachment change or change of mounting orientation.
- 4. Live measurement: with the Peak Switch at OFF the pointer follows the applied force in real time; use this state to observe continuous force values. To lock the current reading, simply switch the Peak Switch to ON.
- 5. Peak hold: with the Peak Switch at ON the pointer stays at the maximum force value of that measurement and does not fall back. To start the next peak measurement, first switch back to OFF (pointer returns to zero), then back to ON.
The mechanical movement has a 10 mm stroke, giving a stable loading travel from contact to full scale; when reading, keep the line of sight perpendicular to the dial to avoid parallax.
V. Key points for digital models (ZTS / ZTA series)
- ZERO: with no load applied, press ZERO to zero the display. Re-zero after changing attachments as well.
- PEAK: press PEAK to enter peak hold mode; the screen locks the maximum force during the process. Press again to return to live display. When running repeated peak tests, reset the peak before each measurement.
- High-speed sampling: ZTS / ZTA sample at 2000 Hz, suitable for capturing true peaks in transient processes such as insertion/extraction and snap-fit forces, without missing spikes as slower sampling would.
- Separate sensor: on ZTS / ZTA the sensor is separate from the display unit, so the sensor can be fixed directly in a fixture or machine while the display is placed where it is easy to read; confirm the cable is properly connected before measuring.
- Capacity coverage: the ZTA series starts at 10 N (1 kgf) and the ZTS series reaches 20 kN (2000 kgf), with resolution stepping up by capacity (e.g. ZTA-LM-10N is 0.01 N / 0.001 kgf, ZTS-LM-20kN is 0.01 kN / 1 kgf).
VI. Main specification comparison (official data for models listed on this site)
| Series | Representative models | Rated capacity | Resolution / graduation | Type |
| PS | PS-30lb | 30 lbf | 0.25 lbf (120 lines) | Mechanical dial |
| PSM | PSM-5K | 5 kgf | 25 gf | High-accuracy mechanical |
| ZTA | ZTA-LM-10N | 10 N(1 kgf) | 0.01 N(0.001 kgf) | Digital, separate sensor |
| ZTS | ZTS-LM-1000N | 1000 N(1 kN) | 1 N(0.1 kgf) | Digital, separate sensor |
| ZTS | ZTS-LM-20kN | 20 kN(2000 kgf) | 0.01 kN(1 kgf) | Digital, separate sensor |
PS series indication accuracy is within ±0.1% F.S., with a 10 mm movement stroke, 8 standard attachments S-1 to S-8, and a carrying case and inspection certificate supplied. The PSM series is the highest-accuracy mechanical grade, covering 20 N to 50 kgf. lbf-graduated models such as the PS-30lb are international models (officially noted as not sold in Japan due to Japanese metrology law); when ordering, make sure the graduation unit matches your actual working practice.
VII. Mounting on a test stand: the stand applies the load, the gauge reads it
- The stand is a loading mechanism: test stands (e.g. SH-1000N horizontal, HV-1000N vertical long-stroke) do not measure force themselves; the force value comes from the fitted force gauge.
- SH-1000N (horizontal): maximum load 1000 N, trapezoidal screw with 3 mm lead, feed 1.2 mm/revolution, maximum sample length approx. 250 mm, net weight approx. 15 kg. Samples lie horizontally for easy clamping, well suited to pull-off tests on terminals and wire harnesses.
- HV-1000N (vertical long-stroke): maximum load 1000 N, same 1.2 mm/revolution feed, maximum sample height approx. 375 mm, stainless steel table, M6 gauge mounting thread, net weight approx. 18 kg; suited to high-elongation or long samples.
- Mounting procedure: 1) Fix the force gauge to the stand mounting position using the M6 thread and confirm it is not skewed; 2) fit the attachment and zero (re-zeroing is mandatory after mounting on the stand); 3) turn the handwheel to apply load at a uniform speed; 4) read the peak once the target force is reached or the sample fails.
- Option notes: SH / HV series can be ordered with an -S length scale or -L stroke +150 mm; compatible gauges include the DST / DSV and ZTS / ZTA series, ordered separately.
VIII. Common mistakes and corrections
- Over-range use: IMADA officially warns against exceeding rated capacity and against impact loading. If the force is unknown, trial-measure with a high-capacity gauge first, then switch to a suitable range.
- Using the wrong measuring shaft: measuring compression on the tension shaft or tension on the compression shaft is explicitly prohibited by IMADA and will damage the measuring shaft.
- Failing to zero with an attachment fitted: forgetting to re-zero after changing an attachment or mounting orientation offsets every reading by the attachment's own weight, which is a very large proportional error at low capacities.
- Side loading: force not along the measuring axis (misaligned sample, misaligned stand guide) gives low readings and damages the shaft assembly. Confirm the sample and probe are coaxial when using a stand.
- Not resetting peak mode: leaving the mechanical Peak switch at ON means the next "peak" is actually a residual value from the previous measurement. The same applies to digital models — clear the peak before consecutive measurements.
- Humid or dusty environments: IMADA notes that high humidity and dust affect accuracy; do not apply oil or chemical soap solution to the measuring shaft, which can cause corrosion or dust build-up.
- Unauthorised disassembly: IMADA warns against removing the warranty seal or dismantling the body; opening the instrument voids the warranty.
IX. Care and storage
- Wipe the body with a soft dry cloth; do not use strong solvents or abrasive cleaners.
- For long periods of non-use, store in the carrying case in a dry place, away from direct sunlight and severe temperature changes.
- Periodic calibration is a prerequisite for maintaining accuracy; the supplied inspection certificate only evidences factory condition. Calibrate on a cycle if the instrument is used frequently or metrological traceability is required.
- During storage, periodically check the measuring shaft and attachments for wear or deformation, and replace any damaged parts immediately.
X. Frequently asked questions
- Q: Mechanical or digital — how do I choose?A: If you only need a force reading, work in an environment with electrical noise or static, and want maintenance-free operation, choose mechanical (PS / PSM); if you need to log data, capture transient peaks or integrate into equipment, choose digital (ZTS / ZTA).
- Q: How do I choose the capacity?A: Select a rated capacity 1.5–2 times the maximum expected force, so normal readings fall within 20%–80% of full scale.
- Q: Are attachments interchangeable?A: IMADA measuring shafts use an M6 thread, so attachments with the same thread are interchangeable; always fit them by hand, without tools.
- Q: Why do readings differ each time?A: First check whether an attachment was fitted without re-zeroing, whether there is side loading, and whether peak mode was reset. Handheld measurement is inherently less repeatable than stand-mounted measurement; use a test stand for important data.