The KTM-150 is a medium-to-low torque model in the KILEWS (Taiwan, China) KTM series, with a range of 1.5–150.0 kgf·cm (0.15–15.0 N·m) and ±0.5% accuracy. It is mainly used to verify the output torque of electric screwdrivers, for daily production-line checks and torque setting management. Its operation is not complicated, but in the chain "set upper/lower limits → zero → fit the test joint → run once until the clutch trips → read the GOOD/NG lamp", a mistake at any step distorts the reading. This article follows the actual operating sequence and explains each step against KILEWS' official KTM-150 operation video; specifications follow the official KILEWS specification table and our product page.

I. First, be clear about what this instrument measures and what it does not

II. Getting to know the panel: keys, indicators and test joints

As the operation video shows, the front of the yellow body carries, from top to bottom, the LCD display, GOOD / NG indicator lamps and the keypad; on top is the measuring seat for mounting the spring test joint (run-down joint). The key layout is as follows:

Supplied as standard: 3 spring test joints: AJ-150 / AJ-50 / AJ-15K, plus a dedicated charger, manual and carrying case. Different joints suit different torque bands and bit sizes; choose according to the torque range of the screwdriver under test and fit it onto the top measuring seat.

III. Four preparations before power-on

  1. Charging: the dedicated charger takes AC100–240V 50/60Hz input and outputs DC12V 400mA; the Ni-MH batteries (1.2V × 4 cells, 1650 mAh) charge in about 3 hours, give about 12 hours of continuous use, and can be recharged over 300 times.Do not measure while charging, as this affects accuracy (see "Precautions" on our product page).
  2. Select and fit the test joint: choose the appropriate AJ-150 / AJ-50 / AJ-15K for the expected torque, seat it firmly with no looseness.
  3. Power-on and zeroing: after power-on confirm the LCD shows 0.00; if a residual reading remains, press CLEAR.
  4. Confirm unit and mode: use UNIT to select the recording unit first, then MODE to select PEAK or TRACK, to avoid discovering after the test that the unit is wrong and the data cannot be compared with the process document.

IV. Setting torque limits (first half of the video)

  1. Press PRG/SET to enter setting mode; the LCD shows the current setting.
  2. Use UP / DOWN to adjust the value and set the lower and upper limits required by the process. The settings should match the process torque tolerance band, not simply push the upper limit to full scale — that would make the GOOD/NG lamps useless as a filter.
  3. Press PRG/SET to confirm and exit setting. From then on every test peak is compared with these limits and the GOOD / NG lamps give a direct pass/fail judgement, with no need to compare numbers manually one by one.
  4. When recording several groups of data, fix the unit and limits first and do not change the settings midway, otherwise earlier and later data are not comparable.

KILEWS describes the function as "a torque range identification LED can be set; users can define the range to be tested, and the torque meter shows a warning lamp if the result does not meet requirements" — the instrument performs the judgement, the operator only sets and reads.

V. Getting one valid reading: the formal test procedure

  1. Alignment: insert the screwdriver bit vertically into the cushion coupling of the spring test joint, keeping it coaxial with the measuring axis and applying no side load; side loading is the most common source of error in torque testing.
  2. Run until trip: press the screwdriver trigger and let the clutch trip naturally (stop when you hear / feel the "click").Do not keep forcing after the trip; the extra force directly raises the reading.
  3. Reading: in PEAK mode the LCD holds the peak — record that value; TRACK mode shows the live value and is useful for watching torque rise.
  4. Judgement: watch the GOOD / NG lamps. Within the set range the green GOOD lamp lights; outside it the red NG lamp warns.
  5. Zero and repeat: press CLEAR to zero before the next test. We recommend 5–10 consecutive tests, then use AVE for the average together with maximum / minimum to judge screwdriver repeatability and clutch stability — one good reading does not mean a stable batch.

The video repeatedly shows power-on zeroing, PRG/SET and UP/DOWN setting, and the GOOD lamp lighting — exactly the "set → zero → test → judge" cycle above.

VI. Full operation video (playable directly)

Below is KILEWS' official KTM-150 product operation video, 4 min 11 s long, demonstrating power-on zeroing, UNIT / PRG/SET / UP / DOWN setting, the auto-zero switch and GOOD / NG judgement. Originally portrait, it has been converted to H.264/AAC web format with a seekable progress bar.

Video 1 KILEWS KTM-150 Torque Tester — official operation demonstration (open in new window / download)

KILEWS KTM-150 torque tester panel and spring test joint

Figure 1 KTM-150 main unit (LCD, GOOD / NG lamps, keypad) and the spring test joint on top

VII. Statistics, memory and USB data output

VIII. Common mistakes and precautions

IX. Choosing within the KTM series: range comparison

The KTM series is graded by range; keep your usual torque in the middle of the range (about 20%–80%). The table below gives the test ranges from the official KILEWS specification table.

Modelkgf·cmN·mTypical application
KTM-60.02~8.00.2~80.0 cN·mMicro torque, precision electronics assembly
KTM-150.15~15.00.015~1.500Low-torque electric screwdrivers
KTM-50S0.5~50.00.05~4.90General medium-to-low torque
KTM-1501.5~150.00.15~15.0Medium-to-low torque electric screwdrivers (the model in this article)
KTM-25050~250.05~25.00Medium-to-high torque screwdrivers
KTM-100050~10005–100 (accuracy graded)High torque, lithium-battery tools

Among them the KTM-150 covers 0.15–15.0 N·m, the most common medium-to-low torque band on production lines. If your usual torque is below 0.15 N·m, look at the KTM-15 or KTM-50S; if it is regularly above 15 N·m, move up to the KTM-250 / KTM-1000.

X. KTM-150 quick specification reference

Test range1.5~150.0 kgf·cm / 0.15~15.0 N·m / 1.5~130.0 Lbf·in
Accuracy±0.5%
Units / modesSwitchable kgf·cm, N·m, Lbf·in; TRACK / PEAK
Power4.8 V DC, Ni-MH rechargeable batteries 1.2 V × 4 cells 1650 mAh; about 3 h charging, about 12 h runtime, over 300 charge cycles
ChargerInput AC100–240V 50/60Hz, output DC12V 400mA
Dimensions / weight230 (W) × 125 (D) × 65 (H) mm, approx. 1.8 kg
Standard accessoriesSpring test joints AJ-150 / AJ-50 / AJ-15K, one each; dedicated charger; manual; carrying case
FunctionsUpper/lower limit LED judgement, auto zero, maximum / minimum / average statistics, USB 1.0 data output

XI. Sources